Occupational radiation monitoring and quality assurance in hospital X-ray departments.

  • Aba A. Bentil Andam
  • C. Schandorf
  • E. K. Addison
Keywords: diagnostic radiology, quality assurance, shielding, workload, radioprotection

Abstract

Quality assurance is essential to any programme of diagnostic radiology, to prevent undue exposure of x-ray personnel and patients to radiation.

This paper reports observations from a survey of some hospital x-ray facilities in Accra and Kumasi in Ghana, to find out about steps taken by x-ray workers to ensure quality control and administration in diagnostic radiology.

The grid point method has been used to do a detailed study of scatter radiation levels of two x-ray departments in Accra.

The iso-dose rate contours have been analysea, and results show the importance of good design of x-ray rooms and provision of protective structures.

Published
2016-01-08
How to Cite
Andam, A. A. B., Schandorf, C., & Addison, E. K. (2016). Occupational radiation monitoring and quality assurance in hospital X-ray departments. Journal of Science and Technology, 10(2), 79- 85. https://doi.org/10.4314/just.v10i2.1076
Section
Articles